Objective: To compare the biometric values measured by Dual Scheimpflug-based and partial coherence interferometry-based optic biometers and to evaluate the intraocular lens (IOL) power calculations performed by using different formulas including SRK-2, SRK-T, Holladay 1 and Hoffer Q. Material and Methods: Patients who applied to our clinic with visual impairment and diagnosed as senil cataract were recruited in the current study. After detailed ophthalmologic examination, biometric measurements including mean anterior chamber depth (ACD), axial length (AL) and keratometric parameters (K1, K2) were performed by using the Galilei G6 Dual Scheimpflug Analyzer and IOL Master 500. The IOL power (Sensar AAB00 hydrophobic acrylic IOL AMO) calculations targeting emmetropia were performed by using SRK-2, SRK-T, Holladay 1 and Hoffer Q formulas with the two devices. These parameters were compared with statistical analysis. Results: One hundred and six eyes of 94 patients with nuclear sclerosis were recruited. The mean age of patients (40 female, 54 male) was 67.44±8.99 years. The value of K1 was found to be statistically higher in measurement performed by IOL Master 500 (p<0.001). The IOL power was calculated statistically higher with all formulas measured by IOL Master 500 (p<0.001). Conclusion: The mean ACD, AL and K2 values did not differ between the two devices. Average K and K1 parameter was measured higher when measured with IOL Master 500. The mean IOL power was calculated statistically significantly different between these two devices.
Keywords: Biometry; intraocular lens implantation; cataract
Amaç: Çift Scheimflug analizatörü ve parsiyel koherens interferometri, optik biometri ile göz içi lensinin (GİL) SRK-2, SRK-T, Holladay- 1 ve Hoffer Q formülleri ile ölçümlerinin karşılaştırılması amaçlandı. Gereç ve Yöntemler: Kliniğimize görme azlığı şikâyeti ile gelen ve senil nükleer katarakt tanısı konulan hastalar çalışmaya dâhil edildi. Hastalara ayrıntılı oftalmolojik muayene sonrası, Gallilei G6 çift Scheimflug analizörü ve İOL Master 500 parsiyel koherens interferometri optik biometri ile ön kamara derinliği (ÖKD), aksiyel uzunluk (AU), keratometrik parametreler (K1, K2), SRK-2, SRK-T, Holladay- 1 ve Hoffer Q formülleri ile GİL ölçümleri (Sensar AAB00 hidrofobik akrilik GİL AMO) postoperatif emetropi hedeflenerek ölçüldü ve istatistiksel analiz yapıldı. Bulgular: Kırk kadın, 54 erkek; 94 senil nükleer katarakt hastasının, 106 gözü çalışmaya alındı. Hastanın yaş ortalaması 67,44±8,99 idi. Ortalama K1 değeri İOL Master 500 parsiyel koherens interferometri optik biometri ile istatistiksel olarak daha yüksek bulundu. (p<0,001). Ortalama GİL ölçüm sonuçları, İOL Master 500 parsiyel koherens interferometri optik biometri ile istatistiksel olarak daha yüksek bulundu. (p<0,001). Sonuç: Ortalama ÖKD, AU ve K2 değerleri bakımından 2 ölçüm arasında fark yoktu. Ortalama K1 değeri İOL Master 500 parsiyel koherens interferometri optik biometri ile istatistiksel olarak daha yüksek bulundu. İki cihaz arasında, göz içi mercek ölçümleri arasında tüm formüllerde farklılık mevcuttur
Anahtar Kelimeler: Biometri; intraoküler lens implantasyonu; katarakt
- Sahin A, Hamrah P. Clinically relevant biometry. Curr Opin Ophthalmol. 2012;23(1):47-53. [Crossref] [PubMed] [PMC]
- Drexler W, Findl O, Menapace R, Rainer G, Vass C, Hitzenberger CK, et al. Partial coherence interferometry: a novel approach to biometry in cataract surgery. Am J Ophthalmol. 1998;126(4):524-34. [Crossref] [PubMed]
- Muzyka-Woźniak M, Oleszko A. Comparison of anterior segment parameters and axial length measurements performed on a Scheimpflug device with biometry function and a reference optical biometer. Int Ophthalmol. 2019;39(5):1115-22. [Crossref] [PubMed]
- Wang ZY, Yang WL, Li DJ, Chen W, Zhao Q, Li YF, et al. Comparision of biometry with the Pentacam AXL, IOL Master 700 and IOL Master 500 in cataract patients. Comparative Study. 2019;55(7):515-21.
- Henriquez MA, Zunigia R, Camino M, Camargo J, Ruiz-Montenegro K, Izquierdo L. Effectiveness and agreement of 3 optical biometers in measuring axial lenght in the eyes of patients with mature cataracts. J Cataract Refractive Surg. 2020 46(9):1222-8. [Crossref]
- Ventura BV, Ventura MC, Wang L, Koch DD, Weikert MP. Comparison of biometry and intraocular lens power calculation performed by a new optical biometry device and a reference biometer. J Cataract Refract Surg. 2017;43(1):74-9. [Crossref] [PubMed]
- Shajari M, Cremonese C, Petermann K, Singh P, Müller M, Kohnen T. Comparison of axial length, corneal curvature, and anterior chamber depth measurements of 2 recently introduced devices to a known biometer. Am J Ophthalmol. 2017;178:58-64. [Crossref] [PubMed]
- Polat O, Baysal Z, Özcan S, İnan S, İnan ÜÜ. Comparison of anterior segment measurements obtained by aladdin optical biometer and sirius corneal topography. Turk J Ophthalmol. 2016;46(6):259-63. [Crossref] [PubMed] [PMC]
- Rabsilber TM, Jepsen C, Auffarth GU, Holzer MP. Intraocular lens power calculation: clinical comparison of 2 optical biometry devices. J Cataract Refract Surg. 2010;36(2):230-4. [Crossref]
- Holzer MP, Mamusa M, Auffarth GU. Accuracy of a new partial coherence interferometry analyser for biometric measurements. Br J Ophthalmol. 2009;93(6):807-10. [Crossref] [PubMed]
- Buckhurst PJ, Wolffsohn JS, Shah S, Naroo SA, Davies LN, Berrow EJ. A new optical low coherence reflectometry device for ocular biometry in cataract patients. Br J Ophthalmol. 2009;93(7):949-53. [Crossref] [PubMed]
- Shin MC, Chung SY, Hwang HS, Han KE. Comparison of two optical biometers. Optom Vis Sci. 2016;93(3):259-65. [Crossref] [PubMed]
- Kongsap P. Comparison of a new optical biometer and a standard biometer in cataract patients. Eye Vis (Lond). 2016;3:27. [Crossref] [PubMed] [PMC]
- Ruiz-Mesa R, Abengózar-Vela A, Ruiz-Santos M. Comparison of a new Scheimpflug imaging combined with partial coherence interferometry biometer and a low-coherence reflectometry biometer. J Cataract Refract Surg. 2017;43(11):1406-12. [Crossref] [PubMed]
- Savini G, Hoffer KJ, Shammas HJ, Aramberri J, Huang J, Barboni P. Accuracy of a new swept-source optical coherence tomography biometer for IOL power calculation and comparison to IOLMaster. J Refract Surg. 2017;33(10):690-5. [Crossref] [PubMed]
- Asena L, Akman A, Güngör SG, Dursun Altınörs D. Comparison of keratometry obtained by a swept source oct-based biometer with a standard optical biometer and scheimpflug imaging. Curr Eye Res. 2018;43(7):882-8. [Crossref] [PubMed]
- Jung S, Chin HS, Kim NR, Lee KW, Jung JW. Comparison of repeatability and agreement between swept-source optical biometry and dual-scheimpflug topography. J Ophthalmol. 2017;2017:1516395. [Crossref] [PubMed] [PMC]
- Gao R, Chen H, Savini G, Miao Y, Wang X, Yang J, et al. Comparison of ocular biometric measurements between a new swept-source optical coherence tomography and a common optical low coherence reflectometry. Sci Rep. 2017;7(1):2484. [Crossref] [PubMed] [PMC]
- Savini G, Hoffer KJ, Schiano-Lomoriello D, Barboni P. Intraocular lens power calculation using a Placido disk-Scheimpflug tomographer in eyes that had previous myopic corneal excimer laser surgery. J Cataract Refract Surg. 2018;44(8):935-41. [Crossref] [PubMed]
.: İşlem Listesi