Objective: We aimed to evaluate the repeatability of iridocorneal angle (IA), anterior chamber depth (ACD), and anterior chamber volume (ACV) measurements of the rotating Scheimpflug camera combined with a Placido disc anterior segment analyser in unoperated and post-refractive surgery eyes. Material and Methods: In this prospective study, we measured IA, ACD, and ACV using a Sirius (CSO, Florence, Italy). Glaucoma summary mode of the Sirius provides nasal and temporal IA values at different angles (specifically; 0°, +10°, +20°, +30°, -10°,-20°, and-30° degrees) automatically and an average value as IA. We calculated repeated measures one-way analysis of variance (ANOVA), intraclass correlation coefficient (ICC), withinsubject standard deviation (Sw), intrasession test-retest variability, and coefficient of variation (COV) to determine repeatability. Results: This study included 50 eyes of 50 subjects without a history of refractive surgery with a mean age of 42.6 years and 13 eyes of 13 subjects who had undergone LASIK or PRK with a mean age of 39.1 years. We did not find any statistically significant differences between consecutive measurements using repeated measures ANOVA in both groups. For the average IA, ACD, and ACV measurements, repeatability was excellent with ICC > 0.90 in both groups. For the IA measurements at different meridians, the ICC value was below 0.9 at all nasal meridians with the exception of nasal angle 10o in operated eyes and higher than 0.8 for all temporal meridians. Conclusion: Iridocorneal angle measurements of the Sirius device at temporal meridians showed moderate or high repeatability, whereas at nasal meridians, it showed poor or moderate repeatability in both normal and post-refractive surgery eyes. On the other hand, ACD and ACV measurements of the Sirius showed excellent repeatability in both groups.
Keywords: Corneal topography; refractive surgical procedures; anterior chamber
Amaç: Bu çalışmada, ameliyat edilmemiş ve refraktif cerrahi geçirmiş gözlerde Placido disk ile bağlantılı rotasyonel Scheimpflug kamera içeren bir ön segment analiz cihazı ile elde edilen iri-dokorneal açı (İA), ön kamara derinliği (ÖKD) ve ön kamara hacmi (ÖKH) ölçümlerinin tekrarlanabilirliğini değerlendirmeyi amaçladık. Gereç ve Yöntemler: Bu prospektif çalışmada, Sirius (CSO, Floransa, İtalya) cihazı kullanarak İA, ÖKD ve ÖKH'yi ölçtük. Sirius'un glokom özet modu, farklı açılarda (spesifik olarak 0°, +10°, +20°, +30°, −10°,−20°, and -30° dereceleri) nazal ve temporal İA değerlerini otomatik olarak ölçer ve ilaveten bir ortalama İA değeri sunar. İA, ÖKD ve ÖKH ölçümleri tekrarlanabilirliği, tek yönlü varyans analizi (ANOVA), gözlem içi korelasyon katsayısı (İKK), bireye özgü standart sapma (Sw), sınıf içi test-tekrar test değişkenliği ve varyasyon katsayısı (VK) hesaplanarak değerlendirildi. Bulgular: Bu çalışmada, ortalama yaşı 42,6 olan ve refraktif cerrahi öyküsü olmayan 50 hastanın 50 gözü ve yaş ortalaması 39,1 yıl olan LASIK veya PRK tedavisi geçirmiş 13 hastanın 13 gözü yer aldı. Her iki grupta ardışık ölçümler, tekrarlanan ölçümler ANOVA testi kullanılarak değerlendirildi ve istatistiksel olarak anlamlı bir fark bulunamadı. Ortalama İA, ÖKD ve ÖKH ölçümleri için her iki grupta da tekrarlanabilirlik mükemmeldi (İKK>0,90 idi). Farklı meridyenlerdeki İA ölçümleri için, İKK değerleri, ameliyat edilen gözlerde nazal 10o değeri hariç tüm nazal meridyenler için 0,9'un altında ve tüm temporal meridyenler için 0,8'in üstünde idi. Sonuç: Sirius cihazının iridokorneal açı ölçümleri, temporal meridyenlerde orta veya yüksek tekrarlanabilirlik gösterirken, nazal meridyenlerde hem normal hem de refraktif cerrahi geçirmiş gözlerde zayıf veya orta derecede tekrarlanabilirlik göstermiştir. Diğer yandan, Sirius'un ÖKD ve ÖKH ölçümleri her iki grupta da mükemmel tekrarlanabilirlik göstermiştir.
Anahtar Kelimeler: Kornea topografisi; refraktif cerrahi işlemler; ön kamara
- Nongpiur ME, Ku JY, Aung T. Angle closure glaucoma: a mechanistic review. Curr Opin Ophthalmol. 2011;22(2):96-101. [Crossref] [PubMed]
- Wu RY, Nongpiur ME, He MG, Sakata LM, Friedman DS, Chan YH, et al. Association of nar-row angles with anterior chamber area and volume measured with anterior segment optical co-herence tomography. Arch Ophthalmol. 2011;129(5):559-74. [Crossref] [PubMed]
- Kaur M, Shaikh F, Falera R, Titiyal JS. Optimizing outcomes with toric intraocular lenses. Indi-an J Ophthalmol. 2017;65(12):1301-13. [Crossref] [PubMed] [PMC]
- Chen X, Miao H, Naidu RK, Wang X, Zhou X. Comparison of early changes in and factors affecting vault following posterior chamber phakic Implantable Collamer Lens Implantation without and with a central hole (ICL V4 and ICL V4c). BMC Ophthalmol. 2016;16(1):161. [Crossref] [PubMed] [PMC]
- Olsen T. Calculation of intraocular lens power: a review. Acta Ophthalmol Scand. 2007;85(5):472-85. [Crossref] [PubMed]
- Fremont AM, Lee PP, Mangione CM, Kapur K, Adams JL, Wickstrom SL, et al. Patterns of care for open-angle glaucoma in managed care. Arch Ophthalmol. 2003;121(6):777-83. [Crossref] [PubMed]
- Fénolland JR, Puech M, Baudouin C, Labbé A. [Imaging of the iridocorneal angle in gloucoma]. J Fr Ophthalmol. 2013;36(4):378-83. [Crossref] [PubMed]
- Konstantopoulos A, Hossain P, Anderson DF. Recent advances in ophthalmic anterior segment imaging: a new era for ophthalmic diagnosis? Br J Ophthalmol. 2007;91(4):551-7. [Crossref] [PubMed] [PMC]
- Nguyen AT, Liu T, Liu J. Applications of scheimpflug imaging in glaucoma management: cur-rent and potential applications. J Ophthalmol. 2016;2016:3062381. [Crossref] [PubMed] [PMC]
- Kurita N, Mayama C, Tomidokoro A, Aihara M, Araie M. Potential of the pentacam in screen-ing for primary angle closure and primary angle closure suspect. J Glaucoma. 2009;18(7):506-12. [Crossref] [PubMed]
- De la Parra-Colín P, Garza-León M, Barrientos-Gutierrez T. Repeatability and comparability of anterior segment biometry obtained by the Sirius and the Pentacam analyzers. Int Ophthalmol. 2014;34(1):27-33. [Crossref] [PubMed]
- Hernández-Camarena JC, Chirinos-Saldaña P, Navas A, Ramirez-Miranda A, de la Mota A, Jimenez-Corona A, et al. Repeatability, reproducibility, and agreement between three different Scheimpflug systems in measuring corneal and anterior segment biometry. J Refract Surg. 2014;30(9):616-21. [Crossref] [PubMed]
- Wang Q, Ding X, Savini G, Chen H, Feng Y, Pan C, et al. Anterior chamber depth measure-ments using Scheimpflug imaging and optical coherence tomography: repeatability, reproduci-bility, and agreement. J Cataract Refract Surg. 2015;41(1):178-85. [Crossref] [PubMed]
- Chen W, McAlinden C, Pesudovs K, Wang Q, Lu F, Feng Y, et al. Scheimpflug-Placido topog-rapher and optical low-coherence reflectometry biometer: repeatability and agreement. J Cata-ract Refract Surg. 2012;38(9):1926-32. [Crossref] [PubMed]
- Masoud M, Livny E, Bahar I. Repeatability and intrasession reproducibility obtained by the sirius anterior segment analysis system. Eye Contact Lens. 2015;41(2):107-10. [Crossref] [PubMed]
- Ruiz-Belda C, Piñero DP, Ruiz-Fortes P, Soto-Negro R, Moya M, Pérez-Cambrodí RJ, et al. Intra-session repeatability of iridocorneal angle measurements provided by a Scheimpflug pho-tography-based system in healthy eyes. Graefes Arch Clin Ophthalmol. 2016;254(1):169-75. [Crossref] [PubMed]
- Prakash G, Srivastava D. Single session, intrauser repeatability of anterior chamber biometric and corneal pachy-volumetric parameters using a new Scheimpflug+Placido device. J Optom. 2016;9(2):85-92. [Crossref] [PubMed] [PMC]
- Savini G, Barboni P, Carbonelli M, Hoffer KJ. Repeatability of automatic measurements by a new Scheimpflug camera combined with Placido topography. J Cataract Refract Surg. 2011;37(10):1809-16. [Crossref] [PubMed]
- Bland JM, Altman DG. Measurement error and correlation coefficients. BMJ. 1996;313(7048):41-2. [Crossref] [PubMed] [PMC]
- Bland JM, Altman DG. Measurement error. BMJ. 1996;313(7059):1654. [Crossref] [PubMed] [PMC]
- Bland JM, Altman DG. Measurement error proportional to the mean. BMJ. 1996;313(7049):106. [Crossref] [PubMed] [PMC]
- Grewal DS, Brar GS, Jain R, Grewal SP. Comparison of Scheimpflug imaging and spectral do-main anterior segment optical coherence tomography for detection of narrow anterior chamber angles. Eye (Lond). 2011;25(5):603-11. [Crossref] [PubMed] [PMC]
- Savini G, Carbonelli M, Barboni P, Hoffer KJ. Repeatability of automatic measurements per-formed by a dual Scheimpflug analyzer in unoperated and post-refractive surgery eyes. J Cata-ract Refract Surg. 2011;37(2):302-9. [Crossref] [PubMed]
.: Process List